Ann Clin Microbiol 2020;23:135-148. Evaluation of the Performance of ASTA MicroIDSys, a Novel Matrix-Assisted Laser Desorption/Ionization-Time of Flight Mass Spectrometry System, in Identification of Bacterial Clinical Isolates

Table 4. Comparison of bacterial identification between ASTA MicroIDSys and Bruker Biotyper for 88 other isolates
Reference identification (number of isolates)ASTA MicroIDSys, number of isolates Bruker Biotyper, number of isolates
Correctly identifiedMis-IDInvalid-ID Correctly identifiedMis-IDInvalid-ID
Genus levelSpecies level Genus levelSpecies level
Candida         
C. albicans (18)1818   1818  
C. tropicalis (16)1616   1616  
C. glabrata (6)66   66  
C. krusei (2)22   22  
C. parapsilosis (2)22   22  
Corynebacterium         
C. striatum (11)1111   1111  
C. tuberculostrearicum (2)22   22  
C. accolens (1)11   11  
Others         
Elizabethkingia meningoseptica (6)66   66  
Saccharomyces cerevisiae (2)22   22  
Actinomyces neuii (1)11   11  
Actinomyces odontolyticus (1)11   11  
Aeromonas caviae (1)11   11  
Bacillus pumilus (1)11   11  
Bacteroides fragilis (1)11   11  
Brevibacterium casei (1)11   11  
Campylobacter jejuni (1)11   11  
Clostridioides difficile (1)11   11  
Clostridium innocuum (1)11   11  
Eggerthella lenta (1)11   11  
Eikenella corrodens (1)11   11  
Finegoldia magna (1)11   11  
Haemophilus influenza (1)11   11  
Leuconostoc lactis (1)11   11  
Neisseria macacae (1)11   10  
Nocardia nova (1)00 1* 00 1
Paenibacillus amylolyticus (1)11   11  
Pichia norvegensis (1)11   11  
Prevotella intermedia (1)11   11  
Propionibacterium avidum (1)11   11  
Rothia mucilaginosa (1)11   11  
Vibrio parahaemolyticus (1)11   11  
Total number of isolates (%)87 (98.9)87 (98.9) 1 (1.1) 87 (98.9)86 (97.7) 1 (1.1)

*Strains not in the database of the instrument.