Ann Clin Microbiol 2021;24:21-26. Comparative Evaluation of the STANDARD F Influenza A/B FIA Test with the Sofia Influenza A+B FIA and SD BIOLINE Influenza Ag A/B/A(H1N1) tests for Influenza A Virus Detection
Test | Sensitivity % (95% CI) | Specificity % (95% CI) |
---|---|---|
STANDARD F Influenza A/B FIA | 50.7 (38.4-63.0) | 100 (97.1-100) |
Sofia Influenza A+B FIA | 50.7 (38.4-63.0) | 100 (97.1-100) |
SD BIOLINE Influenza Ag | 29.9 (19.6-42.4) | 100 (97.1-100) |
Abbreviations: CI, confidence interval; FIA, fluorescent immunoassay.